1: Basics of Evoked Potentials
Evoked potentials are electrical potentials recorded over various parts of the nervous system in response to sensory or motor stimulation. Each stimulation results in a low amplitude evoked response. In order to adequately visualize and measure these responses, evoked potentials are usually averaged and amplified. When many evoked responses have been averaged, they are referred to as an evoked potential. Each evoked potential has a series of waves or peaks in response to the stimulus.
Types of Evoked Potentials
Clinical evoked potentials performed in a clinical laboratory most often involve stimulation of one of the sensory systems and recording the evoked potentials off peripheral nerves, spinal cord, or brain. The motor pathways can also be stimulated with an electrical or magnetic stimulus; however, neither of these is done routinely in an outpatient environment. Electrical stimulation is too painful in an awake patient, and magnetic stimulation is mostly used in research applications. Evoked potentials are named according to the neural pathway that is stimulated.
Modalities
Three types of evoked potentials are routinely used in clinical practice: visual, auditory, and somatosensory. Visual evoked potentials (
Each type of evoked potential consists of several peaks occurring at latencies ranging from a few ms to several hundred ms after stimulation. Peaks are designated as short, middle, or long latency waveforms depending on when they occur. In general, short latency waveforms are clinically more useful as they are more easily reproducible, consistently identified in normal subjects, and resistant to medication effects. Short latency
Nomenclature
The individual peaks of the evoked potentials that are important in clinical evaluation are named according to their latency after stimulation, polarity, or number in a sequence.
Certain peaks are designated as “obligate waveforms.” Obligate waveforms are so named because their absence usually denotes an abnormality. Not all waveforms of an evoked potential are considered obligate. In tibial nerve
Waveform Generators
Evoked potentials are generated when impulses travel through the nervous system. The various components of an evoked potential are generated when there is synaptic transmission or when an impulse is traversing a fiber tract, particularly when that fiber tract changes direction (1). An example of the latter is when a somatosensory impulse ascends from the dorsal horn to the dorsal column pathway. In the cervical region, this transition contributes to the N13 waveform potential seen after median nerve stimulation.
When evoked potentials are recorded close to the generator site, they are referred to as near-field potentials. To record a near-field potential, the recording electrode is placed directly over or very close to the peripheral nerve, spinal cord, or brain structure contributing to the evoked potential. These potentials are often triphasic, with a large negativity preceded and followed by a smaller positivity. The initial positivity occurs due to a wave of depolarization approaching the recording electrode. The large negativity occurs when the depolarization passes beneath the recording electrode, and the final, small positivity occurs due to repolarization. Near-field potentials are of high amplitude and very sensitive to electrode placement. Slight changes in the location of the recording electrode can drastically change the amplitude of the response. The P37 waveform obtained after tibial nerve stimulation is an example of a near-field potential whose amplitude is greatly dependent on the location of the recording electrode (Figure 1.1).
In contradistinction to near-field potentials are far-field potentials. Far-field potentials are recorded from an electrode that is distant to the site of the waveform generator. These potentials are biphasic, with the recording electrode seeing only a moving phase of depolarization through neural pathways. Far-field potentials are of low amplitude and do not significantly change in amplitude and morphology with slight movement of the recording electrodes. The N34 waveform obtained after tibial nerve stimulation is such a waveform (Figure 1.1). Most of the
Evoked potentials can also be thought of as being cortical, subcortical/spinal cord, and peripheral nerve evoked potentials. Cortical evoked potentials are generated by synaptic activity in cortical neurons and the thalamocortical projections. These potentials are usually near-field potentials. The
Specific neural structures are often regarded as the generators of particular evoked potential waveforms. These associations have been made mostly from lesion and autopsy studies. Waveforms have progressively ascending neural structures assigned to them. When a particular waveform is absent or prolonged, the site of pathology is at or distal to its generator. While clinically useful, this approach is too simplistic and has some shortcomings. First, while particular neural generators are thought to result in certain evoked potential waveforms, it should be recognized that a single nucleus or fiber tract may contribute to more than one waveform. Each waveform in turn may be composed of potentials from many different sites (2). Additionally, evoked potentials do not ascend in the nervous system like an electrical signal through a cable, and evoked potential peaks do not simply represent different points on this cable. Evoked potentials may ascend through several different pathways and since many different components contribute to each waveform, the waveforms may appear out of order. An example of this is sometimes seen with median nerve
Age: 14 years
Sex: Female
Stimulation rate: 5.7/s
Filters: 30–3,000 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 8 ms/div
Side: Right
Stimulation duration: 0.2 ms
Stimulation intensity: 18.8 mA
Number of repetitions: 2,000
Tibial nerve conduction velocity (tibial nerve to T12S): 48 m/s
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
8.30 | ||
20.8 | ||
P31 | 27.5 | |
N34 | 30.8 | |
P37 | 34.4 | 2.38 (P37–N45) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
26.1 | |
13.6 |
Discussion: The P37 waveform is a near-field potential whose amplitude is greatly dependent on the location of the active recording electrode. Notice that the P37 waveform is of high amplitude in the
Stimulation Methodology
The type of stimulus used to obtain evoked potentials depends on the type of evoked potential being performed. A light flash or patterned stimulus is used for
Electrodes
Stimulating electrodes are used for
Parameters
Stimulation parameters include intensity, duration, rate, and repetitions. Varying these parameters may result in alteration of the evoked potential waveform morphology and latency.
Intensity
The lowest stimulation intensity that is needed to produce an evoked potential is known as the threshold intensity. As the intensity increases the latency decreases and amplitude increases. The latency and amplitude change is not linearly related to the stimulation intensity, and the changes occur up to a maximum. Beyond this limit, further increase in stimulation intensity does not affect the evoked potentials. Central components of the evoked potentials reach maximum amplitude earlier than peripheral components. This occurs because of “central amplification” of the stimulus (3). Central amplification refers to the ability of a peripheral stimulus to activate more fibers in central pathways than in peripheral nerves. Consequently, with gradually increasing stimulus intensity, cortical and subcortical evoked potentials will cease to become larger at a lower intensity than peripheral evoked potential. This is discussed in more detail in Chapter 4.
Intensity of stimulation is measured differently for different types of evoked potentials. For
Duration
Duration of the stimulus is closely related to the intensity. It refers to the time that the stimulus is applied. Increasing the duration of the stimulus has an effect similar to increasing intensity. Longer duration stimuli will result in shorter latency and higher amplitude of the evoked potential waveforms with the same caveats as noted above for intensity.
Rate
The stimulus rate is the number of stimuli delivered per second. It is often denoted in hertz (Hz), but this is incorrect, as Hz implies that the stimulus is sinusoidal. The stimulus rate must be such that the entire evoked potential waveform of interest is recorded before the next stimulus is delivered. Thus, the rate is dependent on the latency of the evoked potential waveforms. Short latency waveforms require a shorter time window of recording and can be acquired using faster stimulation rates. Long latency potentials require a longer time window of recording and need a slower stimulation rate. For
Age: 56 years
Sex: Female
Stimulation rate: 5.7/s
Filters: 30–1,500 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 3 ms/div
Side: Right
Stimulation duration: 0.2 ms
Stimulation intensity: 16.4 mA
Number of repetitions: 1,213
Median nerve conduction velocity (median nerve to
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.90 | ||
N13 | 13.9 | |
P13 | 13.2 | |
N18 | 16.8 | 2.31 (P13–N18) |
N20 | 19.8 | 1.08 (N20–P22) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
9.90 | |
3.30 | |
P13–N20 | 6.60 |
Discussion: The N13 waveform is thought to arise from the cervical spinal cord and the P13 (and P14) waveform from the nucleus cuneatus. In this example, the P13 waveform has a shorter latency than the N13 waveform even though its generator is rostral to the generator of the N13 waveform. This suggests that individual waveforms cannot be considered to arise from single generators. In reality, each waveform receives contributions from several generators and each generator contributes to multiple waveforms.
In general, with slower rates of stimulation, the evoked potential waveform morphology is better, latency shorter, and amplitude higher. Faster rates stress neural transmission, particularly through synapses, resulting in longer latencies and lower amplitudes. However, the slower the rate, the longer it takes to acquire the data which may compromise patient cooperation. The optimal stimulation rate must be a compromise between evoked potential morphology and speed of acquisition. In some laboratories, evoked potentials will be obtained with a “fast” rate initially. If the evoked potential is normal at this rate, the test is complete. However, if it is abnormal, the evoked potential will be repeated at a slower rate. Occasionally, an evoked potential is normal with the slow rate, but abnormal with fast rate. How such a test is interpreted is controversial. Some would interpret an evoked potential as normal if it is normal at any stimulation rate. On the other hand, some would interpret an evoked potential abnormal if it is abnormal at any stimulation rate. The author’s practice is the former (Figures 1.3A and B).
Repetitions
The number of repetitions refers to the number of evoked responses that must be averaged to produce a reliable evoked potential. The smaller the amplitude of the evoked potential and higher the amplitude of noise (low signal-to-noise ratio), the more repetitions will be needed. The number of repetitions needed is reduced if the signal-to-noise ratio is improved. To some extent this can be done by increasing stimulus intensity and duration and reducing noise. Higher number of repetitions leads to better evoked potential waveform morphology and reproducibility (Figures 1.4A–E). The number of repetitions that are recommended for each type of evoked potentials modality will be discussed in the respective chapters.
Replication
Replication should not be confused with repetitions. Whenever an evoked potential is obtained, it must be reproduced at least once to ensure that the waveforms recorded are reproducible. This is known as replication of the evoked potential. When evoked potentials are not very reproducible, more than one replication may be necessary to confirm the presence of low amplitude waveforms. Most modern evoked potential machines acquire two repetitions simultaneously; that is, alternate stimuli are used to create two different averages.
Recording Methodology
The methods used for recording evoked potentials vary to some degree depending on the evoked potential type. However, there are many aspects of recording that are similar, such as recording electrodes, averaging, and postacquisition signal processing.
Electrodes
Different types of electrodes can be used to record evoked potentials. Any electrode used should have certain properties that will allow reliable data acquisition.
Properties
Recording electrodes must be able to conduct neural signals without distortion. They must be made of material that resists polarization and does not interact with skin or other human tissue with which it makes contact. Electrodes coated with gold, platinum, or silver coated with silver chloride are most often used. Periodically the electrode’s resistance must be checked. This is the opposition to direct current flow and is measured with an ohmmeter. Electrode resistance should be no more than a few ohms, and if it is very high, a breach in the integrity of the electrode should be suspected. Electrode impedance is the opposition to
Age: 49 years
Sex: Male
Filters: 150–3,000 Hz
Ear inserts (add 0.9 ms): Yes
Side: Right
Click polarity: Rarefaction
Stimulation intensity: 70 dBnHL
Masking intensity: 40 dBnHL
Number of repetitions: 2,000
Stimulation rate: 51.1/s
Scale: Amplitude = 0.05 μV/div; Latency = 1 ms/div
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | N/A | N/A (I–In) |
III | 4.88 | |
V | 6.84 | 0.27 (V–Vn) |
Vc | 6.86 | |
N/A (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | N/A |
I–III | N/A |
III–Vc | 1.98 |
Stimulation rate: 11.1/s
Scale: Amplitude = 0.1 μV/div; Latency = 1 ms/div
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 2.40 | 0.07 (I–In) |
III | 4.66 | |
V | 6.64 | 0.35 (V–Vn) |
Vc | 6.72 | |
5.0 (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | 4.32 |
I–III | 2.26 |
III–Vc | 2.06 |
Discussion: Faster stimulation rates can cause degradation in waveform morphology, loss of amplitude, and prolongation of latency. A
Age: 43 years
Sex: Male
Stimulation rate: 2.1/s
Filters: A = 1–100 Hz; B = 10–100 Hz; C = 1–30 Hz
Preauricular–preauricular distance: 38 cm
Scale: Amplitude = 2 μV/div; Latency = 20 ms/div
Eye: Right
Visual angle: 30′
Visual acuity: 20/20
Corrective lenses: Yes
(A)
Number of repetitions: 1
(B)
Number of repetitions: 10
(C)
Number of repetitions: 50
(D)
Number of repetitions: 100
(E)
Number of repetitions: 200
Discussion: This
Types
Surface and needle electrodes can be used to record evoked potential signals. Electroencephalogram (
Needle electrodes can also be used to record evoked potentials. Because they are more invasive than surface electrodes, needle electrodes are not typically used for outpatient evoked potential studies. They can be applied more quickly than surface electrodes as skin preparation is not necessary. This is useful when evoked potentials are performed in the operating room (
Placement
Electrode placement depends on the type of evoked potential being recorded.
Amplifiers
The amplifier is the backbone of any evoked potential recording device. A few basic properties of these amplifiers are presented.
Basics
Amplifiers have the task of taking signals that are of very low amplitude and isolating them from background noise. The signals are then amplified to a degree that can be measured and appropriate calculations are performed. Amplifiers consist of an input board, selector switches, differential amplifier, and filters.
Input board
The input board, sometimes also called the jack box or head box, is a small box that is placed near the patient. Electrodes attached to the patient are plugged into receptacles in this box. The input signals are amplified by the input board before they are transmitted to the main unit of the evoked potential machine. This allows the biologic signal to be amplified before any exogenous noise is introduced as the signal travels through cables. Receptacles in the input board also limit the amount of current that can flow from the machine to the patient, so patients cannot receive electrical shocks from the machine.
Selector switches
Selector switches allow pairing two electrode inputs into the amplifier. This allows the amplifier to compare the signal between the two electrodes. It is with selector switches that recording montages are created.
Differential amplifier
The basic principle of a differential amplifier is that it magnifies and displays the difference between two inputs. Each differential amplifier has two main inputs called Input 1 and Input 2; previously these were also called Grid (G) 1 and G2. Because Input 1 is often close to the site generating the potential of interest, it is also referred to as the Active electrode, whereas Input 2 is referred to as the Reference electrode.
The output of a differential amplifier is the difference between Input 1 (G1, active) and Input 2 (G2, reference) electrodes. Even though one electrode is called “active” and the other “reference,” both contribute to the ultimate waveform. The display convention used determines if the waveforms deflect upward or downward. At times, more than one electrode can be linked to create one reference (Input 2, G2) electrode. An example of this would be linking the two ears as a single reference.
Because the differential amplifier displays the difference between two inputs, signals that are similar in both inputs will be subtracted. This is known as rejection of the common mode signal. An example of signal that is often rejected because it is in common mode is
Averaging
Neurophysiologic signals such as evoked potentials that have amplitude that is lower than the accompanying noise can only be visualized with averaging (5). Averaging involves adding successive responses and dividing the sum by the number of responses. Since the evoked potential is time locked to the stimulus and always occurs at the same latency while noise is random, averaging allows reduction of noise and emergence of the evoked potential signal. Averaging involves digitization of the analog signal in an analog-to-digital (
The number of responses that need to be averaged to resolve an evoked potential waveform depends on the amplitude of the waveform of interest versus the amplitude of noise. This relationship is called the signal to noise ratio. As discussed earlier, the lower the signal to noise ratio, the more responses that need to be averaged. If the signal to noise ratio is increased, either by increasing the amplitude of the signal or reducing the noise, fewer responses need to be averaged. Averaging improves the signal to noise ratio by a factor equal to the square root of the number of repetitions averaged. It is a common misperception that averaging increases the amplitude of the signal, which is incorrect. It reduces the amplitude of the noise.
Filters
Evoked potentials contain waveforms of a limited frequency range. Filters are used to eliminate unwanted waveforms that are below and above the frequencies of interest. By eliminating low and high frequency activity that is not of interest, fewer repetitions are needed to resolve low amplitude evoked potential waveforms. Analog filters are applied to the signal before it is digitized.
Analog filters are divided into low- and high-frequency filters (
Another special type of filter affects only a very narrow frequency band. This is known as a notch filter. Most often this is a 60 Hz filter in the
Digital filters are used after the signal has been digitized. Several types of digital filters are available and one that is used occasionally in clinical evoked potentials is called “smoothing” (1). Smoothing uses a computer algorithm to average three to five consecutive data points to help eliminate noise and “smooth” out waveforms. Smoothing can be used more than once on a waveform. However, it should be used sparingly as it can affect waveform morphology (Figures 1.6A–C). Digital filtering, however, does not cause phase shifts.
Artifact rejection
Artifact rejection is another technique used to help resolve low-amplitude signals such as evoked potentials. Amplifier input voltage is set so that signals of excessively high voltage that are unlikely to be of interest are automatically rejected and not included in the average. This prevents the average from being contaminated from these high-amplitude signals. The limits of artifact rejection should be set so that signals of interest are not affected. Artifact rejection allows a reduction of the number of responses that need to be averaged. However, if the artifact rejection level is set too low, too many responses will be rejected, increasing the time and responses needed to average.
Interpretation
Interpretation of evoked potentials involves a detailed evaluation of the waveforms. This includes determining the presence, morphology, latency, amplitude, and several other features of waveforms. The patient’s evoked potential is compared to normative data to determine if an abnormality is present.
Age: 43 years
Sex: Male
Stimulation rate: 2.1/s
Filters: A = 1–100 Hz; B = 10–100 Hz; C = 1–30 Hz
Preauricular–preauricular distance: 38 cm
Scale: Amplitude = 2 μV/div; Latency = 20 ms/div
Eye: Right
Number of repetitions: 200
Visual angle: 30′
Visual acuity: 20/20
Corrective lenses: Yes
Absolute Latency | |||
---|---|---|---|
Derivation | N75 (ms) | P100 (ms) | N145 (ms) |
A | 84.1 | 112.0 | 158.3 |
B | 81.3 | 95.6 | 128.0 |
C | 88.7 | 114.5 | 160.1 |
Amplitude | |||
---|---|---|---|
Derivation | N75–P100 (μV) | P100–N145 (μV) | Mean (μV) |
A | 4.20 | 8.21 | 6.21 |
B | 2.20 | 5.10 | 3.65 |
C | 5.86 | 8.34 | 7.10 |
Discussion: Filters can affect the shape and latency of evoked potential waveforms. In Figure A, a
Age: 32 years
Sex: Female
Stimulation rate: 31.1/s
Filters: 150–3,000 Hz
Scale: Amplitude = 0.1 μV/div; Latency = 1.5 ms/div
Ear inserts (add 0.9 ms): Yes
Side: Left
Click polarity: Alternating
Stimulation intensity: 85 dBnHL
Masking intensity: 55 dBnHL
Number of repetitions: 4,000
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 2.43 | 0.19 (I–In) |
III | 4.77 | |
V | 6.69 | 0.44 (V–Vn) |
Vc | 6.75 | |
2.3 (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | 4.32 |
I–III | 2.34 |
III–Vc | 1.98 |
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 2.43 | 0.19 (I–In) |
III | 4.77 | |
V | 6.69 | 0.42 (V–Vn) |
Vc | 6.69 | |
2.2 (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | 4.26 |
I–III | 2.34 |
III–Vc | 1.92 |
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 2.28 | 0.20 (I–In) |
III | 4.77 | |
V | 6.66 | 0.39 (V–Vn) |
Vc | 6.69 | |
2.0 (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | 4.41 |
I–III | 2.49 |
III–Vc | 1.92 |
Discussion: Excessive smoothing can distort evoked potential waveforms. Figure A shows a
Peak Identification
The various evoked potential types used in clinical practice have a typical series of peaks that occur at expected latencies. This allows for nomenclature of the peaks, as discussed earlier. Peaks may be labeled according to their polarity and expected latency (i.e., N20 waveform is a negative peak occurring at approximately 20 ms) or their number in a series (i.e., wave III is the third
A few basic principles apply to identification of all waveforms. Responses that are difficult to ascertain because of low amplitude may be improved by increasing the stimulation intensity or duration. This is particularly true for
At times instead of a single peak, two smaller peaks may appear. This is known as a bi-fed waveform. When a bi-fed waveform is seen, it may be suggestive of pathology or may be an artifact of the recording method. An example of the latter can be seen when the P100 waveform is obtained with a midoccipital to mid frontal (
Excessive noise can also make peak identification difficult. An attempt should be made to reduce the noise by having the patient relax or sleep. The noise can be measured by averaging without stimulating. If a flat baseline is not obtained with this method after a standard number of repetitions, the evoked potentials may be difficult to resolve. The number of repetitions obtained can be increased; however the signal to noise ratio increases by a factor of the square root of the number of repetitions. This makes increasingly higher number of repetitions of diminishing utility.
Modern evoked potential equipment automatically tags waveforms based on a preprogrammed algorithm. This works reasonably well when the response is normal. However, when the response is abnormal, the automated peak labeling may be incorrect. Similarly, evoked potential technologists label waveforms they think correspond to waveforms of interest. Experienced technologists are often correct, but even they may be inaccurate with more complex studies (Figure 1.7). It behooves the interpreter to double check that the waveforms are appropriately tagged.
Latency
Latency measurement is the most important measure of evoked potential waveforms. Two main types of latencies can be determined.
Absolute latency
The absolute latency, also referred to as the peak latency, is measured from the start of the stimulus to peak of the waveform of interest. The onset latency is the time from stimulus onset to the start of the waveform. This latter latency measure is not routinely used as the onset of a waveform is much harder to determine than the peak (Figure 1.8).
The absolute latency of a waveform is the time an impulse takes to travel from the point of stimulation to the generator of the waveform of interest. Because the stimulus is applied to end organ, absolute latency measures conduction in the peripheral and central nervous system (or in the case of
Interpeak latency
Interpeak latency (
Amplitude
As with latency measurements, amplitude of waveforms can be measured in several ways. Amplitude abnormalities are not as reliable as latency abnormalities. Evoked potentials are seldom interpreted as abnormal if only amplitude abnormalities exist.
Peak to peak amplitude
Peak to peak amplitude is the most commonly used amplitude measurement. It is a measure of the voltage difference between successive peaks of opposite polarity. The amplitude of the ascending limb, descending limb, or the average of both is determined (Figure 1.9). Whichever method is used, it is important to be consistent within a laboratory and the normative data must have been obtained with the same method.
Age: 18 years
Sex: Male
Stimulation rate: 5.7/s
Filters: 30–3,000 Hz
Scale: Amplitude = 0.2 μV/div; Latency = 8 ms/div
Side: Right
Stimulation duration: 0.2 ms
Stimulation intensity: 30.9 mA
Number of repetitions: 1,503
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
N/A | ||
P21 (31) | 32.0 | |
N24 (34) | 37.1 | |
P27 (37) | 43.9 | 0.18 (P27–N35) |
N35 (45) | 50.4 |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
N/A |
Discussion: Before looking at waveform latencies, the interpreting physician must first evaluate the waveforms and confirm that they have been properly tagged. This is a peroneal nerve
Age: 30 years
Sex: Male
Stimulation rate: 5.7/s
Filters: 30–1,500 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 3 ms/div
Side: Left
Stimulation duration: 0.2 ms
Stimulation intensity: 10.1 mA
Number of repetitions: 750
Median nerve conduction velocity (median nerve to
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.3 | ||
N13 | 10.6 | |
P13 | 12.9 | |
N18 | 16.1 | 1.78 (P13–N18) |
N20 | 18.2 | 1.52 (N20–P22) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
8.90 | |
3.60 | |
P13–N20 | 5.30 |
Discussion: Waveform latencies can be determined either to their onset or peak. Peak latencies are used more often in evoked potentials as onset may be difficult to identify. In this example, the N20 waveform onset latency can be either where the light or dark arrow is placed. The peak latency is where the N20 waveform is tagged. More important than peak (or absolute) latencies are
Baseline to peak amplitude
The baseline to peak amplitude is the voltage from the baseline to the peak of the waveform of interest. The baseline is the immediate post stimulus period of the evoked potential (Figure 1.9). This measurement may be difficult to determine as the baseline may not be horizontal.
Amplitude ratio
The ratio of the peak to peak amplitude of two waveforms of the same evoked potential can be determined and is thought to be a more sensitive amplitude measurement than peak to peak amplitude of a single waveform. This is commonly used in
Age: 39 years
Sex: Female
Stimulation rate: 3.9/s
Filters: 1–100 Hz
Preauricular–preauricular distance: 33 cm
Scale: Amplitude = 5 μV/div; Latency = 25 ms/div
Eye: Right
Number of repetitions: 100
Visual angle: 30′
Visual acuity: 20/25
Corrective lenses: No
Absolute Latency | ||||
---|---|---|---|---|
Derivation | N75 (ms) | P100 (ms) | N105 (ms) | N145 (ms) |
80.0 | 102.0 | 128.0 | ||
80.0 | 101.0 | 126.0 | ||
80.0 | 101.0 | 126.0 | ||
106.0 | ||||
82.5 | 101.0 | 125.0 | ||
82.5 | 103.0 | 127.0 |
Amplitude | |||
---|---|---|---|
Derivation | N75–P100 (μV) | P100–N145 (μV) | Mean (μV) |
12.2 | 16.1 | 14.2 | |
10.5 | 14.6 | 12.6 | |
9.68 | 13.0 | 11.3 | |
4.95 | 8.55 | 6.8 | |
10.7 | 14.2 | 12.5 |
P100 Amplitude Ratio | |
---|---|
Location | Ratio |
1.2 | |
1.8 |
Discussion: Amplitude of a waveform can be measured from baseline to peak or from peak to peak. The latter is used more often in evoked potentials since identifying peaks is more reliable than identifying the baseline. In the example shown, the baseline to peak (narrow line) P100 waveform amplitude is not as reliable as the N75–P100 waveform peak to peak amplitude (broad line). Peak to peak amplitudes can be measured for the descending or ascending limb of the peak, or it can be an average of the two. The above example displays amplitudes of the descending and ascending limb, but it is the average that is used in the author’s laboratory. Amplitude ratios are used in
Other Measurements
It is also important to review other measures that are assessed in the evoked potentials laboratory which are not directly related to the evoked potential. These include measures such as visual acuity for
Variability
Evoked potentials obtained with similar methodology may appear different within the same patient when recorded at different times, and between patients the variability may be even more. There are many reasons for this variability.
Intraindividual variability
Many factors within a patient may affect an evoked potential. Foremost is the attention level. Cortical responses are of higher amplitude when the patient is awake than if they are drowsy or asleep. If the patient is awake when the study is obtained on one side and falls asleep when the other side is acquired, there may be considerable variability of the evoked potential (6). This is seen most remarkably in tibial nerve
Interindividual variability
Evoked potentials can vary considerably between individuals. Sex and body or head size, which are related, can have a significant effect on evoked potential waveform latency. Age can also have an effect on evoked potentials. The biggest change is from infancy to maturity, but changes can also occur with old age. Anatomical variability of neural structures may also cause morphological differences in the evoked potentials. An example of this is the P37 cortical waveform obtained after tibial nerve stimulation. Because the leg somatosensory cortex is near the vertex, the P37 waveform is typically best recorded contralateral to the side of waveform generation (ipsilateral to side of stimulation since dorsal column pathway fibers decussate in the medulla). However, occasionally the P37 waveform is better recorded over the vertex or ipsilateral to site of waveform generation.
Normal Evoked Potentials
When interpreting an evoked potential study, various aspects of the data are compared to normative values. Details of what to evaluate for each modality will be presented in the respective chapters. Normative data are obtained by performing evoked potentials on a group of individuals who do not have neurologic disease. It is important to include both sexes and the spectrum of ages that will be evaluated. At the extremes of ages, it is useful to have separate norms. The data are evaluated to determine if it is normally distributed. If it is, the mean values for latency and amplitude of the waveforms of interest have to be calculated. The upper limit of normal is then defined as the mean value plus 2.5 or 3 standard deviations above the mean. At least 20 individuals should be tested to determine the mean latency and amplitude values. If the data are not normally distributed, the percentile method is used, which indicates the probability that the evoked potential from the control group will be similar to the acquired evoked potential. In the latter method, about 100 subjects must be studied to obtain normative data. Ideally, all evoked potential laboratories should obtain their own normative data by using methodology that is similar to that which will be used for performing clinical studies. However, obtaining normative data is difficult, especially for small laboratories. In these instances, it is acceptable to use published normative data if clinical studies are done using the same methodology as the reference laboratory. Additionally, a few studies on normal individuals should be done using the reference laboratory’s methodology to confirm that normal results are obtained.
Age: 43 years
Sex: Male
Stimulation rate: 5.7/s
Filters: 30–3,000 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 8 ms/div
(A)
Side: Left
Stimulation duration: 0.2 ms
Stimulation intensity: 26.6 mA
Number of repetitions: 2,006
Tibial nerve conduction velocity (tibial nerve to T12S): 41 m/s
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.3 | ||
23.9 | ||
P31 | 32.9 | |
N34 | 35.9 | |
P37 | 40.3 | 1.52 (P37–N45) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
31.0 | |
16.4 |
(B)
Side: Right
Stimulation duration: 0.2 ms
Stimulation intensity: 25.0 mA
Number of repetitions: 3,000
Tibial nerve conduction velocity (tibial nerve to T12S): 41 m/s
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.1 | ||
24.1 | ||
P31 | 32.9 | |
N34 | 34.9 | |
P37 | 40.8 | 2.75 (P37–N45) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
31.7 | |
16.7 |
Discussion: State change of the patient, such as waking up or falling asleep, can affect evoked potential waveforms, particularly tibial nerve
In clinical evoked potentials interpretation, the most important variable to analyze is the latency. For each type of evoked potential, a series of
The amplitude of waveforms should also be determined. As with latency, there are some waveforms for each type of evoked potential for which amplitude is routinely determined. Most laboratories do not use peak to peak or baseline to peak amplitude to determine normalcy of waveform. When amplitude is used, it is often the amplitude ratio that is evaluated. The amplitude ratio may be of different waveforms of the same evoked potential (wave V/I ratio of the
Morphology and topography of the waveform are generally not used in determining whether an evoked potential study is normal. As noted earlier, both of these are highly variable between individuals. Often changing montages and recording from different electrodes to compensate for variable topography will result in more “normal” appearing waveforms.
Abnormal Evoked Potentials
As noted earlier, the most significant evoked potential abnormality is latency prolongation. Amplitude and shape of the waveform should also be evaluated, but these are seldom used as the sole criteria for abnormality. The abnormalities seen in evoked potentials can be used to localize the site of the potential lesion.
Latency prolongation
Prolongation of the latency of a waveform implies that there is slowing of the conduction velocity of the neural pathways from the point of stimulation to the neural generator of the waveform. The slowing is often caused by a lesion along the pathway. When recording evoked potentials, the slowing of interest is that which affects the central nervous system. However, when absolute latencies are evaluated, peripheral nerve conduction is also measured. For example, a delayed N20 waveform absolute latency of the median nerve
Absolute latencies and
Amplitude reduction
A reduction in the amplitude of evoked potential peaks is most often accompanied by latency prolongation. The lower amplitude can be explained by temporal dispersion if conduction is slowed. However, at times amplitude reduction is noted without latency prolongation (Figures 1.13A and B). Whereas this may imply axonal loss in the pathway being tested, it may also be due to technical reasons or evoked potential variability. If reduced amplitude is the only significant finding in an evoked potential study, it is seldom designated as abnormal. Most evoked potential laboratories do not have normative data for amplitude. Amplitude ratios are used more often than absolute amplitude values. They are less susceptible to technical issues and normal evoked potential variability.
Age: 60 years
Sex: Male
Stimulation rate: 5.7/s
Filters: 30–1,500 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 3 ms/div
Side: Left
Stimulation duration: 0.2 ms
Stimulation intensity: 21.9 mA
Number of repetitions: 1,000
Median nerve conduction velocity (median nerve to
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
12.8 | ||
N13 | 16.3 | |
P13 | 17.3 | |
N18 | 20.2 | 1.09 (P13–N18) |
N20 | 22.7 | 1.07 (N20–P22) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
9.90 | |
4.50 | |
P13–N20 | 5.40 |
Discussion: Absolute latencies are a measure of both peripheral and central conduction velocity.
Age: 41 years
Sex: Female
Stimulation rate: 5.7/s
Filters: 30–3,000 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 8 ms/div
(A)
Side: Left
Stimulation duration: 0.3 ms
Stimulation intensity: 24.3 mA
Number of repetitions: 1,500
Tibial nerve conduction velocity (tibial nerve to T12S): 43 m/s
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.9 | ||
24.2 | ||
P31 | 29.9 | |
N34 | 34.0 | |
P37 | 38.6 | 1.45 (P37–N45) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
28.7 | |
14.4 |
(B)
Side: Right
Stimulation duration: 0.3 ms
Stimulation intensity: 29.0 mA
Number of repetitions: 1,500
Tibial nerve conduction velocity (tibial nerve to T12S): 43 m/s
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.4 | ||
24.4 | ||
P31 | 31.2 | |
N34 | 36.0 | |
P37 | 43.0 | 1.49 (P37–N45) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
33.6 | |
18.6 |
Discussion: Absolute and interpeak latencies are compared to normative data to determine if pathology exists. Additionally, latencies should be compared from side to side in the same patient. At times, latencies can be within normal limits, but a significant difference is present between the two sides. If this latency difference is beyond the upper limit of normal, even if the latencies themselves are within normal limits, the study is considered abnormal. In this example, left (A) and right (B) tibial nerve
Age: 56 years
Sex: Female
Stimulation rate: 5.7/s
Filters: 30–1,500 Hz
Scale: Amplitude = 0.5 μV/div; Latency = 3 ms/div
(A)
Side: Left
Stimulation duration: 0.2 ms
Stimulation intensity: 12.5 mA
Number of repetitions: 1,500
Median nerve conduction velocity (median nerve to
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
9.72 | ||
N13 | 14.0 | |
P13 | 13.4 | |
N18 | 17.4 | 1.04 (P13–N18) |
N20 | 20.6 | 0.57 (N20–P22) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
10.88 | |
3.68 | |
P13–N20 | 7.20 |
(B)
Side: Right
Stimulation duration: 0.2 ms
Stimulation intensity: 14.9 mA
Number of repetitions: 1,002
Median nerve conduction velocity (median nerve to
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
10.0 | ||
N13 | 13.6 | |
P13 | 15.1 | |
N18 | 17.7 | 1.33 (P13–N18) |
N20 | 20.9 | 1.41 (N20–P22) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
10.9 | |
5.1 | |
P13–N20 | 5.8 |
Discussion: Low amplitude of evoked potential waveforms is seldom considered an abnormal finding unless latencies are also prolonged. Many patient-related factors (i.e., sleep), technical factors, and pathology can cause an amplitude reduction. In this median nerve
Age: 49 years
Sex: Male
Stimulation rate: 3.9/s
Filters: 1–100 Hz
Preauricular–preauricular distance: 34 cm
Scale: Amplitude = 2 μV/div; Latency = 25 ms/div
(A)
Eye: Right
Number of repetitions: 200
Visual angle: 30′
Visual acuity: 20/50
Corrective lenses: No
Absolute Latency | ||||
---|---|---|---|---|
Derivation | N75 (ms) | P100 (ms) | N105 (ms) | N145 (ms) |
78.0 | 118.0 | 155.0 | ||
78.0 | 118.0 | 155.0 | ||
78.0 | 118.0 | 155.0 | ||
136 | ||||
78.0 | 129.0 | 160.0 | ||
78.0 | 131.0 | 161.0 |
Amplitude | |||
---|---|---|---|
Derivation | N75–P100 (μV) | P100–N145 (μV) | Mean (μV) |
5.97 | 7.26 | 6.62 | |
4.11 | 6.61 | 5.36 | |
3.61 | 5.72 | 4.67 | |
6.91 | 8.23 | 7.57 | |
7.93 | 8.39 | 8.16 |
P100 Waveform Amplitude Ratio | |
---|---|
Location | Ratio |
1.02 | |
1.08 |
(B)
Eye: Right
Number of repetitions: 200
Visual angle: 60′
Visual acuity: 20/50
Corrective lenses: No
Absolute Latency | ||||
---|---|---|---|---|
Derivation | N75 (ms) | P100 (ms) | N105 (ms) | N145 (ms) |
72.5 | 110.0 | 143.0 | ||
73.0 | 109.0 | 139.0 | ||
73.0 | 110.0 | 139.0 | ||
111.0 | ||||
73.5 | 117.0 | 144.0 | ||
71.0 | 112.0 | 146.0 |
Amplitude | |||
---|---|---|---|
Derivation | N75–P100 (μV) | P100–N145 (μV) | Mean (μV) |
7.68 | 5.62 | 6.65 | |
6.46 | 4.48 | 5.47 | |
4.95 | 2.60 | 3.78 | |
5.90 | 3.95 | 4.93 | |
5.35 | 1.21 | 3.28 |
P100 Amplitude Ratio | |
---|---|
Location | Ratio |
1.02 | |
1.50 |
Discussion: The P100 waveform can have an abnormal morphology, and the most common such abnormality is a bi-fed peak, called a W-shaped P100 waveform. This may occur because of technical reasons, patient physiology, or visual pathway problems. In example (A), a W-shaped P100 waveform is noted after right eye stimulation. This patient also had reduced visual acuity (20/50). The
Abnormal waveforms
The shape of the waveform is not used in isolation as a determinant of abnormality. This can be very variable depending on not only technical factors but also patient factors, such as alertness and anatomical variations. One type of abnormal waveform seen commonly is a bi-fed peak. This is most commonly seen with a
Absence of waveforms
Complete loss of all waveforms of an evoked potential may be suggestive of severe pathology in the pathway being tested. Before this determination is made, it is important to confirm that technical problems did not cause this finding. Rechecking all components of the evoked potential machine, including the amplifiers and electrodes is very important. If all waveforms except the first are absent, pathology involving the pathway being tested is much more likely (Figure 1.15). This is because presence of the first waveform documents delivery of adequate stimulation to the neural structures.
Localization
The site of abnormality can be localized in the evoked potentials by determining the pattern of abnormalities. When
Report
As with any other test, the report of the evoked potential study is a crucial part of the entire test. A poorly written or inaccurate report calls into question not only the technical quality of the study but also the validity of interpretation. Every report should start with a “History” section. This should briefly describe the symptoms relevant to the evoked potential study being performed. The question being asked by the referring physician should be noted. A list of medications, especially those affecting the nervous system, should be listed. It is worthwhile for the interpreting physician to be aware of the reimbursable indications for evoked potential studies in their area.
The second paragraph of the report should be the “Report” section. In this section, a brief description of the technique used to obtain the data is described. Common items mentioned in this section include which structures were stimulated, whether unilateral stimulation was used and the stimulation rate. Presence of reproducible waveforms is noted. The significant normal and abnormal parameters (mostly latencies) are listed. All latency and amplitude measurements do not need to be noted. When describing an abnormality, it is best to note the side that was stimulated to produce that abnormality. For example, instead of saying, “A right sided abnormality was noted,” it is more descriptive and accurate to say, “An abnormality was noted after right sided stimulation.” This takes into account that an abnormality seen with right-sided stimulation may imply a left hemispheric lesion. Toward the end of the “Report” section, it is important to describe other important measures noted in the evoked potential laboratory. This includes visual acuity for
Age: 5 years
Sex: Male
Stimulation rate: 11.1/s
Filters: 150–3,000 Hz
Scale: Amplitude = 0.05 μV/div; Latency = 1.5 ms/div
Side: Left
Click polarity: Alternating
Stimulation intensity: 85 dBnHL
Masking intensity: 55 dBnHL
Number of repetitions: 2,223
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 1.61 | 0.11 (I–In) |
III | N/A | |
V | N/A | N/A (V–Vn) |
Vc | N/A | |
N/A (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | N/A |
I–III | N/A |
III–Vc | N/A |
Discussion: Absence of all waveforms of an evoked potential can suggest severe pathology of the pathways involved; however, it is possible that technical problems caused these findings. When the first waveform is present and all others absent, the interpretation can be much more certain that nervous system pathology is accounting for the findings. In this example, only the cochlear microphonic is seen. This indicates that the cochlea was successfully stimulated. Absence of all other waves suggests that there is a severe lesion in the auditory pathway proximal to the cochlea.
The next paragraph is the “Interpretation” and should state whether the test is normal or abnormal. If it is abnormal, the abnormalities are listed in order of significance. Calling a study “borderline” should be avoided. The final paragraph is the “clinical correlation.” In many ways, this is the most important part of the report. In this section, the interpreting physician should determine the localization of the abnormality if present and answer the question posed by the referring physician. In the author’s opinion, noting “clinical correlation required” in this section is redundant and should be avoided. Clinical correlation of the evoked potential findings as they relate to the available history is the job of the interpreting physician. Correlating the evoked potentials findings to the entire history of the patient must be done by the referring physician. This latter “clinical correlation” is implied in all tests and does not need to be explicitly restated in the “clinical correlation” section of the report. Examples of reports of normal
Age: 3 years
Sex: Male
Stimulation rate: 11.1/s
Filters: 150–3,000 Hz
Scale: Amplitude = 0.1 μV/div; Latency = 1.5 ms/div
Ear inserts (add 0.9 ms): Yes
Side: Left
Click polarity: Condensation
Stimulation intensity: 70 dBnHL
Masking intensity: 40 dBnHL
Number of repetitions: 2,223
Absolute Latencies | ||
---|---|---|
Waveform | Latency (ms) | Amplitude (μV) |
I | 2.55 | 0.53 (I–In) |
III | 4.89 | |
V | 8.04 | 0.18 (V–Vn) |
Vc | 8.16 | |
0.34 (V/I ratio) |
Interpeak Latencies | |
---|---|
Waveforms | Latency (ms) |
I–Vc | 5.61 |
I–III | 2.34 |
III–Vc | 3.27 |
Discussion:
(A)
Visual Evoked Potential sample report
HISTORY: This is a XX-year-old patient who is undergoing a visual evoked potential study to evaluate for visual pathway dysfunction. Current medications include: XXXXXX
REPORT: Pattern reversal visual evoked responses were obtained using a XX′ check size visual arc following independent stimulation of the left and right eyes at a rate of X/s. This resulted in high signal to noise ratio with good reproducibility of the waveforms. The P100 waveform absolute latencies were normal at XX ms and XX ms following independent stimulation of left and right eyes, respectively. There was no significant side-toside amplitude asymmetry. Parasagittal P100 waveform derivations did not show any significant latency shifts or amplitude asymmetries. The visual acuity was XX and XX of the left and right eyes, respectively.
INTERPRETATION: This is a normal visual evoked response study.
CLINICAL CORRELATION: This study does not demonstrate an abnormality in the visual pathways.
(B)
Brainstem Auditory Evoked Potential sample report
HISTORY: This is a XX-year-old patient who is undergoing a
REPORT: Brainstem auditory evoked potentials were obtained following independent monaural stimulation with XX dBnHL XXXXX clicks delivered at a rate of XXX/s. All obligate waveforms were obtained with good reproducibility. The waves I–Vc
INTERPRETATION: This is a normal
CLINICAL CORRELATION: This study does not demonstrate abnormality in central conduction involving the brainstem auditory pathways.
(C)
Somatosensory Evoked Potential sample report
HISTORY: This is a XX-year-old patient who is undergoing an
REPORT:
INTERPRETATION: This is a normal somatosensory evoked potential study following independent stimulation of bilateral median and tibial nerves.
CLINICAL CORRELATION: There is no evidence of conduction abnormalities along the peripheral nerves stimulated, dorsal columns, or medial lemniscal pathways.
Discussion: These are sample
Conclusions
Clinical evoked potentials are performed in many neurophysiology laboratories, with
References
- .Misulis KE. Spehlmann’s Evoked Potential Primer. 2nd ed. Boston, MA: Butterworth-Heinemann; 1994.
- .Chiappa KH. Principles of evoked potentials. In: Chiappa KH, ed. Evoked Potentials in Clinical Medicine. 3rd ed. Philadelphia, PA: Lippincott-Raven Publishers; 1997:1-30.
- .Eisen A, Purves S, Hoirch M. Central nervous system amplification: its potential in the diagnosis of early multiple sclerosis. Neurology. 1982;32:359-364.
- .ACNS. Guideline 9A: guidelines on evoked potentials. J Clin Neurophysiol. 2006;23:125-137.
- .Chiappa KH, Ropper AH. Evoked potentials in clinical medicine (first of two parts). N Engl J Med. 1982;306:1140-1150.
- .Yamada T, Kameyama S, Fuchigami Y, et al. Changes of short latency somatosensory evoked potential in sleep. Electroencephalogr Clin Neurophysiol. 1988;70:126-136. Figure 1.4 (A–E) Visual evoked potentials (pattern reversal)